- A calibration may not be performed when using an external sensor without color support.
Please refer to Using sensor models without color measurement capability in 7.2.6 Sensor
Setup (page 77). - Additional considerations for details.
- The PICTURE MODE in the ON-SCREEN-DISPLAY (OSD) of some display models will be
changed from its factory preset mode to “GCMDQA” after a calibration. There are cases that
this factory preset status will not be overwritten. However this will not influence the actual
calibration.
- A Black Luminance Measurement may fail due to poor condition of an external sensor (i.e.
due to aging). Please try a different sensor unit. A failed display sensor may be readjusted
using Reference Calibration (page 75), by using a good external sensor.
7.2.2. Rearrange Displays
Click Rearrange Displays in Calibration to rearrange display. See 7.1.1 Rearrange Displays
(page 51).
7.2.3. Calibration Reports
Click Calibration Reports in Display Overview to display the Calibration Reports Display
Selection dialog box, as shown in Figure 34.
Figure 34: Calibration Reports Display Selection dialog box
The Summary tab will be displayed when the OK button is clicked. Click the White
Luminance (Figure 36), Grayscale (Figure 37) or Display Function (Figure 38) tabs to display
the corresponding dialog boxes. The OK button cannot be clicked if no display is selected.
Common settings for each tab
Execution Date If the listbox under Execution Date is clicked, the latest report will
be displayed at the top, with previous reports underneath in order
of date and time. Click the date you wish to view and that day’s
measurement report will be displayed.