Graph
JND per DDL Shows a graph of the JND interval per DDL calculated from the
luminance characteristics after calibration in the conformance test.
Contrast Response Shows a contrast response graph calculated from the luminance
characteristics after calibration in the conformance test. The
contrast response graph is a logarithmic base 10 representation.
JND per DDL The graph shows JND per DDL for a selected Execution Date.
Average and linear fit lines are shown or hidden by checking or
un-checking the Show Average and Show Linear Fit boxes.
Contrast Response This graph shows the contrast response for a selected Execution
Date. The ideal DICOM curve, as well as 10% limit curves and 20%
limit curves are shown or hidden by checking or un-checking the
Show ideal curve, Show 10% Limits and Show 20% Limits boxes.
Buttons Click and drag to draw a rectangle around a certain section, then
use these buttons to zoom in or zoom out within the graph.